    Model 360

    ← Back to products

    Product Overview

    Five element (Na, K, Li, Ca, Ba)

    Single channel Analogue Flame Photometer

    Ergonomic design with front loading, unobstructed sample work area, with removable tray for easy cleaning

    With ease of use resulting from auto ignition and flame optimisation.

    Peak picker for presenting unambiguous readings

    With selected filter indicator

    Safety features including fail safe flame detection and shut off, with low air pressure indicator

    High sensitivity with fine manual control of air, and stable flame design

    And ease of maintenance and storage

    Separate pneumatic and electronic pods

    Small footprint (20 x 30cm)

    Easy to remove and clean filter stick

    Available In Three Formats:

    • 36000009 – M360 5-ELEMENT FLAME PHOTOMETER – Propane/Butane
    • 36000109 – M360 5-ELEMENT FLAME PHOTOMETER – Natural Gas
    • Both the above are supplied with individual1000 ppm (mg/l) calibration concentrate solutions
    • 36000209 – M360C 5-ELEMENT FLAME PHOTOMETER – Propane/Butane
    • The M360C is provided with a mixed calibration concentrate 140.0/5.00/1.50 mmol/l Na/K/Li

    It is important to order the correct part number to ensure fuel gas compatibility. Detector response in the 360C is “linearized” when the Sodium filter is selected; 
the 360C specification – see below – assumes samples are diluted as indicated.

    Model 360C Specification

    Measurement Range

    • Na Serum: 110 to 170 mmol/l
    • Na Urine 0 to 200 mmol/l
    • K Serum 0 to 10.0 mmol/l
    • K Urine 0 to 200 mmol/l
    • Li Serum 0 to 3.0 mmol/l

    Linearity

    • Na Serum: within ± 2 mmol/l
    • Na Urine within ± 4 mmol/l
    • K Serum ‹ ± 0.2 mmol/l
    • K Urine ‹ ± 2 mmol/l
    • Li Serum ‹ ± 0.2 mmol/l

    Reproducibility

    • Na (140 mmol/l) better than 1.5% CV
    • K (5 mmol/l) better than 1.5% CV
    • K (80 mmol/l) better than 1.5% CV
    • Li better than 2.5% CV

    Dilution Factor

    • Na Serum/Urine: 1 in 200 with de-ionised water
    • K Serum: in 200 with de-ionised water
    • K Urine: 1 in 1000 with de-ionised water
    • Li Serum 1 in 50 with de-ionised water

    Note: The above is correct for use with Propane or Butane. Results obtained from 20 replicates of the same sample, aspirating sample for 20 seconds, then water for 10 seconds. To achieve the stated specification the flame must be alight for a minimum of 30 minutes, with diluent being aspirated. (For a brand new, newly installed unit, the stabilisation is highly likely to be greater than 30 minutes).

    Note: For Li determinations; blank and standard solutions must contain similar levels of Na, K and Ca to those present in physiological samples.

    Propane/Butane Specification

    Model 360 Part Number 360 00 009

    OPERATING RANGE

    Sodium is highly prone to self-absorbance in the flame above a concentration of 10 ppm.

    RANGE

    Displayed for all elements – 0 – 1999

    Minimum Detection Level
    @ 3 x Noise
    Concentration
    to give min. 100 units
    Linearity
    % of value at 50% of value
    ppm mg/l meq/l ppm mg/l meq/l
    Na 0.1 0.1 0.0044 1 1 0.0435 ±2% @ 3 ppm
    K 0.1 0.1 0.0025 0.5 0.5 0.0127 ±2% @ 3 ppm
    Li 0.1 0.1 0.0144 1 1 0.1441 ±2% @ 3 ppm
    Ca 2* 2* 0.0998* 20* 20* 0.9980* ±2% @ 100 ppm
    Ba 20* 20* 0.2912* 300* 300* 4.368* ±2% @ 1000 ppm

    *may not be achievable if fuel gas used is Natural Gas (Part Number 360 001 09)

    For optimum performance the Model 360 Flame Photometer should be warmed up for 30 minutes.

    SPECIFICITY

    Na, K & Li interference is <0.5% value of element under analysis.

    REPRODUCIBILITY

    <1.5%* CV for 10ppm K set to 1000 (20 aliquots sampled for 20 seconds each from a single sample). Reading taken 20 seconds after introduction and with a Blank aspirated for 10 seconds between each standard reading.
    <1.0%* CV for 20 consecutive readings taken at 20 second intervals without removing the uptake tube from a 10ppm Potassium solution Set to 1000.

    (*may not be achievable if fuel gas used is Natural Gas)

    ASPIRATION RATE

    3-6 ml/min

    SPEED OF ANALYSIS

    Time to stability 20 seconds

    DATA OUTPUT

    Analogue: Voltage 0.20 Volts FSD

    PACKING DETAILS

    (H x W x D)
    500 mm x 300 mm x 230 mm 5.6 Kg Net
    480 mm x 420 mm x 350 mm 8.6 Kg Packed