Search for...

Or try a general search...

    CLOSE

    Model 410

    ← Back to products

    Product Overview

    All three Models of 410 Flame Photometer come with a preloaded filter stick which houses a maximum of three element filters. The filter stick is easily removed and replaced within the chimney housing (no tools required) and as such it is a relatively simple operation to introduce alternative element specific filters. Filters available for use with all Model 410’s include:

    Sodium, Potassium, Lithium, Calcium, Barium, Strontium, Rubidium and Caesium; the most extensive range available from any Flame Photometer Manufacturer.

    The Model 410 Classic Flame Photometer (410 00 009) comes with Sodium, Potassium and Calcium filters. The “Classic” design is, in essence, the one developed by Corning some 30 years ago. We still manufacture this highly successful unit due to popular demand. The Model 410 Classic Flame Photometer has Sodium, Potassium and Calcium NIST traceable 1000 ppm standard concentrates for each of those analytes (6 x 100 ml). We have expanded the Model 410 Classic’s utility by introduction of the:

    Models 410 Industrial and 410 Scientific Flame Photometers; both come with a Digital Interface fitted as standard. The Digital Interface allows connection to a serial printer or computer for result collection.

    The Model 410 Industrial Flame Photometer (410 00 209)

    Flame Photometry is the technique of choice for the measurement of Sodium, Potassium and Calcium in all sample types in: Mineral Extraction, Oil Industry, Paper Industry, Pharmaceuticals, Soil Analysis, Utilities, Food & Beverage, Chemical Manufacture and Fertilisers. The Model 410 Industrial Flame Photometer with digital output, for printer or PC connection, also has Sodium linearisation capability to allow higher concentrations of Sodium (to 20 ppm) to be directly measured using a single point calibration. That facility can be switched on or off according to the operator’s preference or protocol for Sodium determinations. The Model 410 Industrial Flame Photometer has Sodium, Potassium and Calcium specific filters and NIST traceable 1000 ppm standard concentrates for each of those analytes (6 x 100 ml).

    Model 410 Scientific Flame Photometer (410 00 309)

    For biological samples, Flame Photometry is the most sensitive and robust method for the determination of Sodium, Potassium and Lithium. The Model 410C Flame Photometer is delivered with Na, K and Li filters pre-fitted and an appropriate multi-element calibrator in mmol/l. (1 x 100ml). The Model 410 Scientific instrument is also fitted with a Digital Interface and Sodium linearisation facility allowing direct, straight line, calibration of clinically significant concentrations of Sodium, as well as Potassium and Lithium, following suitable dilution. (Sherwood’s Model 806 Continuous Flow Diluter, used together with the Model 410 Scientific Flame Photometer, gives nominal sample dilution ratios of 1:200 and 1:50. It has been designed to give consistent dilution of calibrator and samples to the Flame Photometer).

    Both Model 410 Industrial and Model 410 Scientific Flame Photometers can be further enhanced by the use of Sherwood’s Model 410 BlueNotes software (410 66 000) and Model 865 Autosampler (865 00 009). Fully integrated and automated systems are possible, even for the Model 410 Classic Flame Photometer via our Update and Automate Modules and packages. See BlueNotes 410 Software in Flame Photometer Accessories on main Flame Photometer product page.

    M410 Classic

    OPERATING RANGE

    Sodium is highly prone to self- absorbance in the flame above a concentration of 10 ppm. The Digital Interface/Lineariser Module accessory (see above) will extend the linear response to 20ppm.

    RANGE Displayed

    Na 0 – 1999

    K 0 – 1999

    Ca 0 – 1999

     

    MINIMUM DETECTION LEVEL

    @3 x noise

    SAMPLE CONCENTRATION

    to give min. 100 Units and still be linear (not Sr)

    LINEARITY

    % of full value at 50% of value

    ppm mg/l meq/l ppm mg/l meq/l
    Na 0.02 0.02 0.0009 0.5 –10 0.5 –10 0.02 – 0.45 ± 1% @ 3 ppm 
    K 0.02 0.02 0.0005 0.5 – 20 0.5 – 20 0.01 – 0.5 ± 1% @ 3 ppm 
    Ca 0.2 0.2 0.003 10 – 100 10 – 100 0.125 – 1.25 ± 1% @ 100 ppm 
    Li 0.02 0.02 0.003 0.5 – 20 0.5 – 20 0.07 – 2.88 ± 1% @ 3 ppm 
    Ba 20 20 0.07 1000 1000 3.65 ± 1% @ 1000 ppm
    Cs 0.2 0.2 0.002 20 – 100 20 – 100 0.15 – 0.75 ± 5% @ 100 ppm*
    Rb 0.1 0.1 0.001 1 – 100 1 – 100 0.01 – 1.17 ± 1% @ 100 ppm
    Sr 20 20 0.1 1000 1000 5.71 Nonlinear > 200 ppm

    * for best results construct a calibration curve for the concentration range of Cs of interest.

    SPECIFICITY

    Na, K & Li Interference is < 0.5% value of element under analysis.

    REPRODUCIBILITY 

    (K=10ppm 20 aliquots from single sample)

    < 1.5%

    ASPIRATION RATE

    3-6 ml/min

    SPEED OF ANALYSIS 

    Time to stability 15 seconds

    DATA OUTPUT

    Analogue: Voltage 0.20 Volts FSD

    DIMENSIONS

    510mm x 390mm x 345mm 9.5 Kg Nett

    730mm x 450mm x 450mm 17 Kg Packed

    M410 Industrial Specification

     

    M410 Industrial Specification

    OPERATING RANGE

    Sodium is highly prone to self- absorbance in the flame above a concentration of 10ppm.The Printer Lineariser Module accessory fitted to the M410 Industrial extends the linear response to 20ppm.

    RANGE Displayed

    Na 0 – 1999

    K 0 – 1999

    Ca 0 – 1999

    MINIMUM DETECTION LEVEL

    @3 x noise

    SAMPLE CONCENTRATION

    to give min. 100 Units and still be linear

    LINEARITY

    % of Nominal value when measured at 50% Nominal value

    Na 20 ppb 0.5 – 10 ppm ** ± 1% @ 3 ppm 
    K 20 ppb 0.5 – 20 ppm ± 1% @ 3 ppm 
    Ca 0.2 ppm 10 – 100 ppm ± 1% @ 100 ppm 
    Li 20 ppb 0.5 – 20 ppm ± 1% @ 3 ppm 
    Ba 20 ppm 1000 ppm ± 1% @ 1000 ppm
    Cs 0.2 ppm 20 – 100 ppm ± 5% @ 100 ppm*
    Rb 0.1 ppm 1 – 100 ppm ± 1% @ 100 ppm
    Sr 20 ppm 1000 ppm Non linear > 200 ppm

    * for best results construct a calibration curve for the concentration range of Cs of interest.

    ** 20ppm with Lineariser switched on.

    SPECIFICITY

    Na, K & Li Interference is < 0.5% value of element under analysis.

    REPRODUCIBILITY 

    (K=10ppm 20 aliquots from single sample)

    < 1.5%

    ASPIRATION RATE

    3-6 ml/min

    SPEED OF ANALYSIS 

    Time to stability 15 seconds

    DATA OUTPUT

    Analogue: Voltage 0.20 Volts FSD

    DIMENSIONS

    510mm x 390mm x 345mm 9.5 Kg Nett

    730mm x 450mm x 450mm 17 Kg Packed

     

    M410 Scientific Specification

    OPERATING RANGE

    Sodium is highly prone to self- absorbance in the flame above a concentration of 10 ppm. A Lineariser is fitted to the Model 410C for clinical applications where normal sodium concentrations even after dilution at 200:1 exceed this concentration. Potassium and Lithium do not exhibit self- absorbance.

    RANGE Displayed

    Na 0 – 1999

    K 0 – 1999

    Li 0 – 1999

    REPRODUCIBILITY 

    (K=10ppm 20 aliquots from single sample)

    < 2%

    SAMPLE CONCENTRATION RANGE

    (before dilution)

    NB Dilution for 

    Clinical Samples

    Linearity Reproducibility

    Cv

    Na   Serum  110 – 170 mmol/l 1:200 ± 2 mmol/l < 1.5%
    Na   Urine       0 – 200 mmol/l 1:200 ± 4 mmol/l < 1.5%
    K    Serum      0 – 10.0 mmol/l 1:200 ± 0.2 mmol/l < 1.5%
    K   Urine         0 – 200 mmol/l 1:5 then 1:200 ± 2.0 mmol/l < 1.5%
    Li                     0 – 3.0 mmol/l 1:50 ± 0.2 mmol/l < 2.5%

    </div”>

    SPECIFICITY

    Na, K & Li Interference is < 1.0 mmol/l with normal levels of other elements in Serum/urine.

    ASPIRATION RATE

    3-6 ml/min

    SPEED OF ANALYSIS

    Time to stability 15 seconds

    DATA OUTPUT

    Analogue: Voltage 0.20 Volts FSD

    DIMENSIONS

    510mm x 390mm x 345mm 9.5 Kg Nett

    730mm x 450mm x 450mm 17 Kg Packed